This page includes support information and frequently asked questions for TestVonics ADC-2500 Series Air Data Calibrators.
If you have any additional questions or need help troubleshooting a problem, please contact our service team.

TTU-205 Series

Vacuum and Pressure Tubing Notice for ADC-2500 Series Air Data Calibrator

This is a notice to all TestVonics Air Data Calibrator users. We have had several calls from users who are finding a Track loss during Auto Exercise. TestVonics found several users using 1/4” tube with an adapter fitting(s) from the vacuum source to the rear panel and getting a track loss condition. For this reason we are releasing this notice for all users to check the following:

  • Vacuum Source Hoses - The hosing from the vacuum source to the ADC rear panel must be at least 3/8” diameter. TestVonics recommends MIL-H-5593-6 hosing to meet this requirement. TestVonics offers a Vacuum Hose Assembly (Part No. ADC-2500-VAC-1-08) for purchase that meets this requirement.
  • Pressure Source Hoses - The hosing going from the pressure source to the ADC-2500 rear panel must be at least 1/4” diameter. TestVonics recommends using MIL-H-5593-4 hosing from the Pressure source. TestVonics offers a Pressure Hose Assembly (Part No. ADC-2500-PRES-1-08) for purchase that meets this requirement.

TestVonics has Vacuum and Pressure Sources available for purchase if an adequate source is not available.

Shipping Instructions for ADC-2500 Series Air Data Calibrator

If the product must be shipped for any reason through a common carrier, it must be packaged properly to minimize the risk of damage. FAILURE TO PROPERLY SHIP THE ADC-2500 SERIES COULD VOID THE SYSTEM WARRANTY. Follow these instructions to properly pack the system for shipment.

What is the best way to power off the ADC-2500 Calibrators?

The ADC-2500 operates on a custom embedded operating platform. Most ADC-2500 systems do not have a built-in write filter to protect the OS from sudden shut downs. All ADC-2522, ADC-2550, and ADC-2555 systems feature write protection, but should still be powered down as instructed below. The safest way to power down the ADC-2500 Series is as follows:
  • Locate the 'Start' button in the lower left corner
  • Select the 'Shut Down' option from the menu
  • They system will shut off, when the screen turns white, switch off the Power
ADC-2500 Part Number
P/N: ADC-2500
NSN: 6695-01-581-8231
Software Version: 3.1.29
EccADC Version: 2017 (USAF Approved)
EccADC Version: 2019 (Commercial)

ADC-2550 Part Number
P/N: ADC-2550
NSN: 6695-01-590-1161
Software Version: 3.1.29

ADC-2522 Part Number
P/N: ADC-2522
NSN: awaiting
Software Version: 3.1.29

ADTS-1555 Part Number
P/N: ADTS-1555
NSN: awaiting
Software Version: 3.1.29

ADC-2500 Series Manuals
ADC-2500 Operating Instructions
Publication No.: TVCS-2500-OPI
Current Revision: 5
USAF T.O. 33D7-45-123-1

ADC-2500 Calibration Instructions
Publication No.: TVCS-2500-CAL
Current Revision: 3

ADC-2500 Illustrated Parts Breakdown
Publication No.: TVCS-2500-IPB
Current Revision: 5
USAF T.O. 33D7-45-123-4

ADC-2500 Remote Communications Manual
Publication No.: TVCS-2500-COM
Current Revision: 4
USAF T.O. 33D7-45-123-8-1

ADC-2550 Series Manuals
ADC-2550 Operating Instructions
Publication No.: TVCS-2550-OPI
Current Revision: 4

ADC-2550 Calibration Instructions
Publication No.: TVCS-2550-CAL
Current Revision: 5

ADC-2550 Illustrated Parts Breakdown
Publication No.: TVCS-2550-IPB
Current Revision: 2

ADC-2550 Remote Communications Manual
Publication No.: TVCS-2500-COM
Current Revision: 4
USAF T.O. 33D7-45-123-8-1